Scientific Analytical Calorimeter Solutions
 

Printing | CAL3K Features

All units have a RS232 printer port. The baud speed can be set from the keyboard. 8 printout formats are available for showing different data sets on the 80 column printer.

Bulk printouts are initiated from the keyboard. Printing can be exclude faulty determinations, or the can by restricted to a GID (Group Identification) or a particular vessel number.

If enabled, the result is printed automatically after a determination.

The printing refers to a ‘paper readable’ output within 80 characters.

The RS232 output can be connected to a PC.

Then a suitable data capturing program such as ‘TERATERM’ can be used to transfer the data to a spreadsheet if so desired.

RS232 outputs with more than 80 characters per line are called ‘DATA DUMPS’

See: DATA

START A CALIBRATION

The calibration of the CAL3K is a process in which a substance (Benzoic Acid, BA) of known calorific value is burned so that future burning of unknown substances can be compared to it. In other words the calibration records all the system variables such as temperature rise and heat loss in memory. All this data is called a calibration field. The calibrations are stored in the vessel.

Learn More

PERFORMING A CALIBRATION

To perform a standard calibration using Benzoic Acid Tablets you will need : A working oxygen bomb calorimeter system connected to a power source, A CAL3K Manual Oxygen Filing Station connected to an oxygen supply (not needed when using the CAL3K-AP Calorimeter)...

Learn More

INTELCAL SYSTEM CALIBRATION

The CAL3K system uses two calibrations: Keyboard calibration, in real time, one sample, average with last ‘n’ calibrations, PC calibration, in retrospect, multiple samples. The PC calibration is called ‘IntelCal’ because of its complexity and system effectiveness.

Learn More